51
III.RS232CommunicationsParameters:
JP5PINS7,8,and9
CommunicationsParameters:
A.BaudRate115,200
B.NoParity
C.1StartBit1StopBit
NOTEifyouareusingcommunicationssoftwaredifferentfromtheonesuppliedwiththecontroller:The
demandsoftemperaturecontrolrequirearelatively
largeportionoftheprocessingpoweroftheonboard
microcontroller.Itispossibleforahostcomputertosenddatatooquicklyforthecontrollertoreceiveand
update.Ifcommunicationsisdisruptedaredisruptedafterrepeatedqueries,TETechnologyrecommends
addingadelaybetweeneachcommandsenttoallow
thecontrollersufficienttimetoprocessandrespond
tothequery.
IV.SerialCommands:
1.MODELCODE
WriteCommand:NA
ReadCommand: 00
Interpret: 9613returned.
2.CONTROLSENSORTEMPERATURE
WriteCommand:NA
ReadCommand: 01
Interpret: Convertreturnedvaluetodecimalanddivideby10.0
10
for
temperaturein°C.
3.POWEROUTPUT
WriteCommand:NA
ReadCommand: 02
Interpret: 511
10
represent100%output.
0
10
returnedis0%output.
4.SECONDARYSENSORTEMPERATURE
WriteCommand:NA
ReadCommand: 04
Interpret: Convertreturnedvaluetodecimalanddivideby10.0
10
for
temperaturein°C
Comments to this Manuals