20
2.0 DisplayandMenuOptions
Note:thissectionappliesonlytotheTC‐48‐20.Setup,programming,andcommunicationswiththeTC‐48‐20OEM
controllercanonlybeaccomplishedbyusingthecomputersoftwarewiththeoptionalMP‐3023RS232cable.
ChangesmadetothecontrollerusingthekeypadarealwaysstoredintheEEPROM
regardlessofEEPROMWRITE
ENABLEsetting.
2.1 Initializationscreen
1.Displaysfor1.5seconds: TETECHNOLOGY
www.tetech.com
2.Displaysfor1.5seconds: TETECHNOLOGY
TC‐48‐20Rev.****
2.2 Primaryscreen
TEMP=###.#°C
OUT=###%T2=##
Thecontrollerdisplaysinformationaboutthefollowingitems:
1)Thetemperatureofthecontrolsensor(TEMP)toa0.1°Cresolution
2)Theoutputlevel(dutycycle)tothethermoelectricmodules(OUT)
3)Thetemperatureofthesecondtemperaturesensor(T2)
Ifthecontrolsensoriseitheranopencircuitorashortcircuit(indicatingthatthesensorisnot
connectedorhasfailed),thecontrollerwilldisplaySENSOR1ERRORonthetoplineofthedisplay
insteadofTEMP=###.#°C.Theoutput%willalsobeforced
to0%untiltheerrorisresolved.
Ifasecondarysensorisnotconnectedtothecontroller,T2=###willnotbedisplayed(unlessitis
forcedtodisplaybyusingthesoftwaremenusetting).
IfALARM1orALARM2hascausedthepoweroutputtobeturned
off,thewordsALARMandOUT=0%will
flashinalternatingsequenceintheplaceofthenormalOUT=###%indicator.Whennotusingathermistor
onthesecondsensorinputtheuserwillneedtoadjusttheALARM2FUNCTIONtoKEEPOUTPUTON.
ParametersMenu
WhenatthePrimaryScreen(shownbelow),depressingtheMENUkeyallowstheusertoscrollthroughand
adjustthevariouscontrollerparameters.
MENU
BUTTON
DOWN ARROW BUTTON
DECREASE PARAMETER
UP ARROW BUTTON
INCREASE PARAMETER
Comments to this Manuals